E8257D/67D, E8663D PSG Signal Generators Service Guide
Assembly Replacement
A25 67 GHz Highband Detector (Options 550 and 567)
3-68
A25 67 GHz Highband Detector (Options 550 and 567)
Tools Required
T–10 driver
5/16” open–ended wrench
Removal Procedure
Refer to Figure 3- 43 for this procedure.
1. Disconnect the power cord.
2. Remove the covers from the signal generator. Refer to “Outer Instrument Cover” on page 3- 4 and “Inner
Instrument Cover” on page 3- 6.
3. Position the signal generator with the A31 Motherboard facing away from you and the RF deck on top.
4. Using the 5/16” open–ended wrench, disconnect the A25 67 GHz Highband Detector from the A24 67 GHz
Highband Coupler.
5. Using the needle–nose pliers, disconnect W15 (not shown) from the A25 67 GHz Highband Detector.
6. Remove the A25 67 GHz Highband Detector.
Replacement Procedure
1. Reverse the order of the removal procedure.
2. Torque all T–10 screws to 9 in–lbs.
3. Torque all RF connectors to 8 in–lbs.
4. Perform the post–repair adjustments and performance tests that pertain to this removal procedure.

Table of Contents