E8257D/67D, E8663D PSG Signal Generators Service Guide
Assembly Replacement
A44 Scan Modulation Bias Adapter (E8257D-520 and E8663D Only)
3-120
A44 Scan Modulation Bias Adapter (E8257D-520 and E8663D Only)
Tools Required
needle- nose pliers
5/16” open–ended wrench
T- 10 driver
Removal Procedure
Refer to Figure 3- 71 for this procedure.
1. Disconnect the power cord.
2. Remove the covers from the signal generator. Refer to “Outer Instrument Cover” on page 3- 4 and
“Inner Instrument Cover” on page 3- 6.
3. Position the signal generator with the A31 Motherboard facing away from you and the RF deck on top.
4. Using the needle- nose pliers, disconnect W148 from A44J2.
5. Using the 5/16” open–ended wrench, disconnect the following RF cables:
W149 from the A44 assembly
W150 from AM MOD BIAS
6. Using the T–10 driver, remove the two screws (1) that attach the A44 Scan Modulation Bias Adapter and bracket
to the side panel.
7. Disconnect the W151 ribbon cable from A26J33.
8. Remove the A44 Scan Modulation Bias Adapter from the RF deck.
Replacement Procedure
1. Reverse the order of the removal procedure.
2. Perform the post–repair adjustments and performance tests that pertain to this removal procedure.

Table of Contents