E8257D/67D, E8663D PSG Signal Generators Service Guide
Assembly Replacement
A35 3–20 GHz I/Q Modulator (E8267D Only - Standard or with Option H16/016) (Serial Prefixes >=US4722/MY4722)
3-102
A35 3–20 GHz I/Q Modulator (E8267D Only - Standard or with Option H16/016) (Serial Prefixes
>=US4722/MY4722)
Tools Required
T–10 driver
5/16” open–ended wrench
needle–nose pliers
Removal Procedure
Refer to Figure 3- 61 for this procedure.
1. Disconnect the power cord.
2. Remove the covers from the signal generator. Refer to “Outer Instrument Cover” on page 3- 4 and “Inner
Instrument Cover” on page 3- 6.
3. Position the signal generator with the A31 Motherboard facing you and the RF deck on top.
4. Using the 5/16” open–ended wrench, disconnect the following RF cables:
W77 from A35J1
W134 from A35 I/Q Mod
5. Using the needle–nose pliers, disconnect the following flexible cables:
W62 from A35J7
W63 from A35J6
W127 from A35J109
W128 from A35J107
W129 from A35J108
W130 from A35J106
6. Using the T–10 driver, remove the three screws (1) that attach the A35 I/Q MOD to the RF deck.
7. Disconnect the W70 ribbon cable from the A35 I/Q MOD.
8. Remove the A35 I/Q MOD from the RF deck.
Replacement Procedure
Reverse the order of the removal procedure. (See the internal cover overlay for cable color codes.)
Torque all T–10 screws to 9 in–lbs.
Torque all RF connectors to 8 in–lbs.
Perform the post–repair adjustments and performance tests that pertain to this removal procedure.