E8257D/67D, E8663D PSG Signal Generators Service Guide
Assembly Replacement
A36 Quadraplier (Options 550 and 567)
3-104
A36 Quadraplier (Options 550 and 567)
Tools Required
T–10 driver
5/16” open–ended wrench
Removal Procedure
Refer to Figure 3- 62 for this procedure.
1. Disconnect the power cord.
2. Remove the covers from the signal generator. Refer to “Outer Instrument Cover” on page 3- 4 and “Inner
Instrument Cover” on page 3- 6.
3. Position the signal generator with the A31 Motherboard facing away from you and the RF deck on top.
4. Using the needle–nose pliers, disconnect the following flexible cables:
W108 from A36J4
W109 from A36J3
W110 from A36J6
W111 from A36J7
W96 from A36J5
5. Using the 5/16” open–ended wrench, disconnect the following RF cables:
W106 from A36J1
W107 from A36J2
W112 from A36J3/RF OUT
6. Using the T–10 driver, remove the four screws that attach the A36 Quadraplier board to the side panel.
7. Disconnect the W113 ribbon cable from the A36J10.
8. Remove the A36 Quadraplier from the RF deck.
Replacement Procedure
1. Reverse the order of the removal procedure.
2. Torque all T–10 screws to 9 in–lbs.
3. Torque all RF connectors to 8 in–lbs.
4. Perform the post–repair adjustments and performance tests that pertain to this removal procedure.