EasyManua.ls Logo

Keysight E8257D/67D - A17 Baseband Generator Interface (E8267 D Only)

Keysight E8257D/67D
344 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
E8257D/67D, E8663D PSG Signal Generators Service Guide
Assembly Replacement
A17 Baseband Generator Interface (E8267D Only)
3-38
A17 Baseband Generator Interface (E8267D Only)
Tools Required
5.0 mm hex–nut driver
9/16” hex–nut driver
needle–nose pliers
Removal Procedure
Refer to Figure 3- 21 for this procedure.
1. Disconnect the power cord.
2. Remove the covers from the signal generator. Refer to “Outer Instrument Cover” on page 3- 4 and “Inner
Instrument Cover” on page 3- 6.
3. Remove the rear panel. Refer to “Rear Panel (Standard)” on page 3- 94 or
“Rear Panel (Option 1EM)” on page 3- 96.
4. Disconnect the W64 ribbon cable from A17P2.
5. Disconnect the W90 ribbon cable from A17P1.
6. Using the needle–nose pliers, disconnect the W68 wire cable from A17P4.
7. Using the 9/16” hex–nut driver, remove the nuts and washers securing the four BNC
connectors (1) on the rear panel.
8. Using the 5.0 mm driver, remove the hex screws (2) and washers that secure the A17 Baseband Generator
Interface to the rear panel.
9. Remove the A17 Baseband Generator Interface.
Replacement Procedure
1. Reverse the order of the removal procedure.
2. Torque all hex screws to 9 in–lbs.
3. Perform the post–repair adjustments and performance tests that pertain to this removal procedure.