EasyManua.ls Logo

Keysight E8257D/67D - A26 MID (Microcircuit Interface Deck)

Keysight E8257D/67D
344 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
E8257D/67D, E8663D PSG Signal Generators Service Guide
Assembly Replacement
A26 MID (Microcircuit Interface Deck)
3-72
A26 MID (Microcircuit Interface Deck)
Tools Required
T–10 driver
Removal Procedure
Refer to Figure 3- 45 and Figure 3- 46 for this procedure.
1. Disconnect the power cord.
2. Remove the covers from the signal generator. Refer to “Outer Instrument Cover” on page 3- 4 and “Inner
Instrument Cover” on page 3- 6.
3. Position the signal generator with the A31 Motherboard facing you and the RF deck on top.
4. If you have an Option 532, 540, or 544 signal generator, disconnect the W29 ribbon cable from A26J33.
5. Disconnect the following ribbon cables:
W69 from A26J14
W28 from A26J15
W19 from A26J30
W36 from A26J13
W33 from A26J32
W34 from A26J31
W35 from A26J10
W20 from A26J20
W113 from A26J34 (E8257D & E8663D only)
W70 from A26J34 (E8267D only)
W99 from the A26J35 (Option 532 and 544 only)
W116 from the A26J36 (Option 532 and 544 only)
6. Disconnect the A25BW1 power cable from A26J12.
7. Using the T–10 driver, loosen the screw (1) that secures the A26 MID to the RF deck.
8. Slide the A26 MID towards the front panel until the locking standoffs are free.
9. Remove the A26 MID from the RF bracket.
Replacement Procedure
1. Reverse the order of the removal procedure.
2. Torque all T–10 screws to 9 in–lbs.
3. Perform the post–repair adjustments and performance tests that pertain to this removal procedure.