EasyManua.ls Logo

Keysight E8257D/67D - Troubleshooting Performance Test Problems; A27 40 GHz Doubler (Option 540)

Keysight E8257D/67D
344 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
E8257D/67D, E8663D PSG Signal Generators Service Guide
Troubleshooting
Troubleshooting Assembly–Level Problems
1-60
2400 Power Supply Test
If the +15V and –15V power supplies are good in the above table, replace the A38 Lowband Filter.
2401 RF Path Test
1. Remove the A38 Lowband Filter and insert a semi–rigid extender cable into J3.
2. Connect the extender cable to a spectrum analyzer.
3. Tune the spectrum analyzer and signal generator to 2 GHz. Turn on the signal generator’s RF.
4. You should see a high power signal (>+20 dBm) at 2 GHz on the spectrum analyzer.
If the signal is good, replace the A38 Lowband Filter.
If the signal is troubleshoot the RF path.
2402 Lowband Amp/Filter Power Supply
Replace the A38 Lowband Filter.
2403 RF Path
Replace the A38 Lowband Filter.
Self–Tests 25xx: A43 Amp Filter Self–Test Errors
+5.2 P142–2, 27 +5.04 +5.36 Main Supply
–5.2 P142–1, 26 –5.30 –5.10 YIG Driver
+5.2 Digital High P141–64, 65, 129, 130 +5.04 +5.36 Main Supply
+3.4 Digital Low P141–60, 61, 62, 63,
125, 126, 127, 128
+3.3 +3.5 Main Supply
+2.6 P141–57, 58, 59, 123,
124
+2.52 +2.68 Main Supply
+1.8 P141–55, 56, 120, 121 –1.75 –1.85 Motherboard
Supply Voltage
(Vdc)
Connector Pins Minimum Value
(Vdc)
Maximum Value
(Vdc)
Origin

Table of Contents