E8257D/67D, E8663D PSG Signal Generators Service Guide
Troubleshooting
Troubleshooting Assembly–Level Problems
1-61
Before proceeding to the reported self–test error code, check the following voltages. If any voltages are out of
specification troubleshoot the supply problem first.
2500 Power Supply Test
If the power supplies, in the table above, are good, replace the A43 Amp Filter.
2501 RF Amp Test
Replace the A43 Amp Filter.
2502 Filter Path Test
Replace the A43 Amp Filter.
2503 Pulse Modulator Test
Replace the A43 Amp Filter.
2504 Step Attenuator Test
Replace the A43 Amp Filter.
2505 Divider On/Off Test
Replace the A43 Amp Filter.
Supply Voltage
(Vdc)
Connector Pins Minimum Value
(Vdc)
Maximum Value
(Vdc)
Origin
+15 P142–28 +14.55 +15.45 Main Supply
–15 P142–3 –15.45 –14.55 Main Supply
+5.2 P142–2, 27 +5.04 +5.36 Main Supply
–5.2 P142–1, 26 –5.30 –5.10 YIG Driver
+5.2 Digital High P141–64, 65, 129, 130 +5.04 +5.36 Main Supply
+3.4 Digital Low P141–60, 61, 62, 63,
125, 126, 127, 128
+3.3 +3.5 Main Supply
+10 P142, 10, 11, 36 +9.7 +10.3 Main Supply

Table of Contents