E8257D/67D, E8663D PSG Signal Generators Service Guide
Troubleshooting
Troubleshooting Assembly–Level Problems
1-36
408 ALC Reference DAC Test
Replace the A8 Output.
409 ALC Modulation Test
Replace the A8 Output.
410 Burst Mod Driver Test
Replace the A8 Output.
411 ALC Detector Test
Replace the A8 Output.
412 Het Band Test
Measure the 1 GHz signal from the A7 Reference board:
1. Turn on the signal generator.
2. Set CW = 200 MHz.
3. Using a spectrum analyzer, probe P51–19.
There should be a 1 GHz signal, that is >30 dBm.
If the signal is good, replace the A9 Output.
If the signal is not good, run the A7 self–tests.