Quick Reference 2
Keysight InfiniiVision 1000 X-Series Oscilloscopes User's Guide 81
Whether reference waveforms are on. Displayed reference waveforms consume
acquisition memory.
Whether segmented memory (available on DSOX1000-Series models) is on.
Acquisition memory is divided by the number of segments.
The horizontal time/div (sweep speed) setting. At faster settings, fewer data
points appear on the display.
When saving to a CSV format file, the maximum number of data points is
50,000.
When necessary, the Length control performs a "1 of n" decimation of the data .
For example: if the Length is set to 1000, and you are displaying a record that is
5000 data points in length, four of each five data points will be decimated,
creating an output file 1000 data points in length.