EasyManua.ls Logo

Keysight InfiniiVision DSOX1102G - Environmental Conditions; Declaration of Conformity

Keysight InfiniiVision DSOX1102G
96 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Quick Reference 2
Keysight InfiniiVision 1000 X-Series Oscilloscopes User's Guide 87
Environmental Conditions
Declaration of Conformity
For Declarations of Conformity for Keysight products, go to:
www.keysight.com/go/conformity
Environment Indoor use only.
Ambient
temperature
Operating: 0 °C to +50 °C
Non-operating: -40 °C to +71 °C
Humidity Operating: 50% to 95% RH at 40 °C for 5 days.
Non-operating: 90% RH at 65 °C for 24 hours.
Altitude Operating: to 3,000 m (9842 ft)
Non-operating to 4,000 m (13,123 ft)
Overvoltage
Category
This product is intended to be powered by MAINS that comply to Overvoltage
Category II, which is typical of cord-and-plug connected equipment.
Pollution Degree The InfiniiVision 1000 X-Series oscilloscopes may be operated in environments
of Pollution Degree 2 (or Pollution Degree 1).
Pollution Degree
Definitions
Pollution Degree 1: No pollution or only dry, non-conductive pollution occurs.
The pollution has no influence. Example: A clean room or climate controlled
office environment.
Pollution Degree 2. Normally only dry non-conductive pollution occurs.
Occasionally a temporary conductivity caused by condensation may occur.
Example: General indoor environment.
Pollution Degree 3: Conductive pollution occurs, or dry, non-conductive
pollution occurs which becomes conductive due to condensation which is
expected. Example: Sheltered outdoor environment.

Table of Contents

Related product manuals