Parameter Typical Return
STEP | RAMP | STAircase | SINusoid | PULSe | TRAPezoid | EXPonential
| UDEFined | CDWell | SEQuence |NONE, *RST NONE
STEP, RAMP, STA, SIN, PULS, TRAP, EXP,
or UDEF.
Specify a step on channel 1: ARB:FUNC:SHAP STEP, (@1)
l Select the CURRent:MODE Arb parameter to enable the instrument to generate arbitrary current
waveforms when the transient system is initiated and triggered.
l Select the VOLTage:MODE Arb parameter to enable the instrument to generate arbitrary voltage
waveforms when the transient system is initiated and triggered.
l Select the POWer:MODE Arb parameter to enable the instrument to generate arbitrary power
waveforms when the transient system is initiated and triggered.
l Select the RESISTANCE:MODE Arb parameter to enable the instrument to generate arbitrary res-
istance waveforms when the transient system is initiated and triggered.
[SOURce:]ARB:FUNCtion:TYPE <type>, (@<chanlist>)
[SOURce:]ARB:FUNCtion:TYPE? (@<chanlist>)
This command replaces the previous ARB:FUNCtion command and should be used in
new applications. ARB:FUNCtion is still available for backward compatibility..
Selects a current, voltage, power, or resistance Arb. Only one type of Arb may be output at a time.
Parameter Typical Return
CURRent | VOLTage | POWer | RESISTANCE *RST VOLTage <type>
Specify an Arb type on channel 1: ARB:FUNC:TYPE CURR, (@1)
l
Power and resistance Arb types apply to .
[SOURce:]ARB:TERMinate:LAST <Bool>, (@<chanlist>)
[SOURce:]ARB:TERMinate:LAST? (@<chanlist>)
Determines the output value when the Arb terminates. The state is either ON (1) or OFF (0).
When ON The output remains at the last Arb sequence value. The last current, voltage, power, or resistance Arb value
becomes the IMMediate value when the ARB completes.
When OFF The output returns to the settings in effectbefore the Arb started. This also applies when the Arb is aborted,
Parameter Typical Return
0|OFF|1|ON, *RST Off 0or1
Sets the ARB to terminate with the voltage, resistance, or current setatthe last Arb value: ARB:TERM:LAST ON,(@1)
l The query returns 0 if the output returns to the settings that were in effect before the Arb started,
and 1 if the output remains at the last Arb value.
Keysight N6700C Operating and Service Guide 187
5 SCPI Programming Reference