Grouped modules, those that are connected in parallel and have been configured or grouped to act as
a single higher-power channel, are assigned the channel number of the lowest numbered slot of the
group. Refer to Output Groups for more information.
Rack Mounting
You cannot use support rails for rack mounting your instrument.
Support rails would block the airflow needed for cooling.
Use Rack Mount kit (Option 908) to rack mount your instrument.
The Rack Mount Kit is also available by ordering part number N6709A.
Keysight N6700 MPS mainframes can be mounted in a 19-inch EIA rack cabinet. They are designed to
fit in one rack-unit (1U) of space. Do not block the air intake and exhaust at the sides of the unit, or the
exhaust at the rear of the unit.
Tools required: Phillips driver, T22 Torx driver, T10 Torx driver
Step 1. Install eight clip-nuts on the rack frame (2 in each corner) where your instrument will be
located.
The following figure references the installation steps.
2 Installation
54 Keysight N6700C Operating and Service Guide