Measure Commands
Measure commands measure the output voltage, current, or power. The MEASure queries start a new
measurement immediately. They are not synchronized to any trigger event.
Use the INITiate, TRIGger, and FETCh commands if a synchronized measurement is necessary.
The default time interval and number of points settings yield a measurement time of 21 milliseconds
per reading (1024 data points at 20.48 microsecond intervals). The default windowing function is
Rectangular.
MEASure[:SCALar]:CURRent[:DC]? (@<chanlist>)
MEASure[:SCALar]:VOLTage[:DC]? (@<chanlist>)
MEASure[:SCALar]:POWer[:DC]? (@<chanlist>)
Initiates and triggers a measurement. Returns the averaged output measurement. Values returned
are either in amperes, volts, or watts. The reading is in the form +n.nnnnnnE+nn for each channel
specified. Multiple responses are separated by commas.
Parameter Typical Return
(none) <DC value>
Returns the averaged current MEAS:CURR? (@1)
Returns the averaged power MEAS:POW? (@1)
Returns the averaged voltage MEAS:VOLT? (@1)
MEASure[:SCALar]:CURRent:ACDC? (@<chanlist>)
MEASure[:SCALar]:VOLTage:ACDC? (@<chanlist>)
Initiates, triggers, and returns the total RMS measurement (AC + DC). Values returned are either in
amperes, or volts. The reading is in the form +n.nnnnnnE+nn for each channel specified. Multiple
responses are separated by commas.
Parameter Typical Return
(none) <ACDC value>
Returns the measured RMS current MEAS:CURR:ACDC? (@1)
Returns the measured RMS voltage MEAS:VOLT:ACDC? (@1)
MEASure[:SCALar]:CURRent:HIGH? (@<chanlist>)
MEASure[:SCALar]:VOLTage:HIGH? (@<chanlist>)
Initiates, triggers, and returns the High level of a pulse waveform. Values returned are either in
amperes, or volts.See Measurement Types. The reading is in the form +n.nnnnnnE+nn for each
channel specified. Multiple responses are separated by commas.