Parameter Typical Return
(none) <HIGH value>
Returns the measured high level current MEAS:CURR:HIGH? (@1)
Returns the measured high level voltage MEAS:VOLT:HIGH? (@1)
MEASure[:SCALar]:CURRent:LOW? (@<chanlist>)
MEASure[:SCALar]:VOLTage:LOW? (@<chanlist>)
Initiates, triggers, and returns the Low level of a pulse waveform. Values returned are either in
amperes, or volts.See Measurement Types. The reading is in the form +n.nnnnnnE+nn for each
channel specified. Multiple responses are separated by commas.
Parameter Typical Return
(none) <LOWvalue>
Returns the measured low level current MEAS:CURR:LOW? (@1)
Returns the measured low level voltage MEAS:VOLT:LOW? (@1)
MEASure[:SCALar]:CURRent:MAXimum? (@<chanlist>)
MEASure[:SCALar]:VOLTage:MAXimum? (@<chanlist>)
MEASure[:SCALar]:POWer:MAXimum? (@<chanlist>)
MEASure[:SCALar]:CURRent:MINimum? (@<chanlist>)
MEASure[:SCALar]:VOLTage:MINimum? (@<chanlist>)
MEASure[:SCALar]:POWer:MINimum? (@<chanlist>)
Initiates, triggers, and returns the maximum or minimum values of a measurement. Values returned
are either in amperes, volts, or Watts. The reading is in the form +n.nnnnnnE+nn for each channel
specified. Multiple responses are separated by commas.
Parameter Typical Return
(none) <MIN value>
<MAX value>
Returns the measured minimum current MEAS:CURR:MIN? (@1)
Returns the measured minimum voltage MEAS:VOLT:MIN? (@1)
Returns the measured minimum power MEAS:POW:MIN? (@1)
Returns the measured maximum current MEAS:CURR:MAX? (@1)
Returns the measured maximum voltage MEAS:VOLT:MAX? (@1)
Returns the measured maximum power MEAS:POW:MAX? (@1)
5 SCPI Programming Reference
262 Keysight N6700C Operating and Service Guide