Front Panel Menu Reference SCPI Command
Select Transient\Arb\Config\Trap\Level.
Enter the Start and Top level in volts. Then press Select.
Select Transient\Arb\Config\Trap\Time.
Enter the Start, Rise, Top, Fall, and End times in seconds..
Then press Select.
To configure a voltage trapezoid Arb, use:
ARB:VOLT:TRAP:STAR 0, (@1)
ARB:VOLT:TRAP:TOP 10,(@1)
ARB:VOLT:TRAP:STAR:TIM 0.25,(@1)
ARB:VOLT:TRAP:RTIM:TIM 0.5,(@1)
ARB:VOLT:TRAP:TOP:TIM 0.5,(@1)
ARB:VOLT:TRAP:FTIM:TIM 0.5,(@1)
ARB:VOLT:TRAP:END:TIM 0.25,(@1)
Step 3. Configure the final steps common to all Arbs. Refer to Configuring Steps Common to all Arbs at
the end of this section.
Configuring Staircase Arbs
Step 1. Select the Arb Type and Shape. Power and Resistance types apply to .
Front Panel Menu Reference SCPI Command
Select Transient\Arb\Function.
Select a Voltage, Current, Power, or Resistance Arb Type.
Then selectthe Staircase Shape. Then press Select.
To selecta current, voltage, power, or resistance staircase
Arb, use:
ARB:FUNC:TYPE VOLT|CURR|POW|RES,(@1)
ARB:FUNC:SHAP STAIR,(@1)
Step 2. Configure the staircase parameters. The parameters apply to the selected Arb type (Voltage,
Current, Power, or Resistance).
Enter the values for the level before and after the staircase.
Enter the time before the staircase, the stair-time of the ramp, and the time after the staircase.
Then enter the total number of steps.
Front Panel Menu Reference SCPI Command
Select Transient\Arb\Config\Stair\Level.
Enter the Start and End level in volts. Then press Select.
Select Transient\Arb\Config\Stair\Time.
Enter the Start, Stair, and End times in seconds.
Enter the number of steps. Then press Select.
To configure a voltage staircase Arb, use:
ARB:VOLT:STA:STAR 0,(@1)
ARB:VOLT:STA:END 10,(@1)
ARB:VOLT:STA:STAR:TIM 0.25,(@1)
ARB:VOLT:STA:TIM 0.5,(@1)
ARB:VOLT:STA:END:TIM 0.25,(@1)
ARB:VOLT:STA:NST 6,(@1)
Keysight N6700C Operating and Service Guide 123
4 Operating the Power System