[SOURce:]ARB:SEQuence:STEP:CURRent:STAircase:NSTeps <value>, <step#>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:CURRent:STAircase:NSTeps? [MIN|MAX,] <step#>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:VOLTage:STAircase:NSTeps <value>, <step#>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:VOLTage:STAircase:NSTeps? [MIN|MAX,] <step#>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:POWer:STAircase:NSTeps <value>, <step#>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:POWer:STAircase:NSTeps? [MIN|MAX,] <step#>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:RESISTANCE:STAircase:NSTeps <value>, <step#>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:RESISTANCE:STAircase:NSTeps? [MIN|MAX,] <step#>, (@<chanlist>)
Specifies the number of steps in the staircase for Arb sequence steps.
Parameter Typical Return
0 – 500 | MIN | MAX, *RST 10 <number of steps>
Step 0 - 99, *RST 0 (none)
Programs five steps for the current staircase: ARB:CURR:STA:NST5,(@1)
Programs five steps for the voltage staircase: ARB:VOLT:STA:NST5,(@1)
Programs five steps for the power staircase: ARB:POW:STA:NST 5,(@1)
Programs five steps for the resistance staircase: ARB:RESISTANCE:STA:NST5,(@1)
Programs five steps for sequence step 0: ARB:SEQ:STEP::CURR:STA:NST 5, 0(@1)
[SOURce:]ARB:CURRent:STAircase:STARt[:LEVel] <value>, (@<chanlist>)
[SOURce:]ARB:CURRent:STAircase:STARt[:LEVel]? [MIN|MAX,] (@<chanlist>)
[SOURce:]ARB:VOLTage:STAircase:STARt[:LEVel] <value>, (@<chanlist>)
[SOURce:]ARB:VOLTage:STAircase:STARt[:LEVel]? [MIN|MAX,] (@<chanlist>)
[SOURce:]ARB:POWer:STAircase:STARt[:LEVel] <value>, (@<chanlist>)
[SOURce:]ARB:POWer:STAircase:STARt[:LEVel]? [MIN|MAX,] (@<chanlist>)
[SOURce:]ARB:RESISTANCE:STAircase:STARt[:LEVel] <value>, (@<chanlist>)
[SOURce:]ARB:RESISTANCE:STAircase:STARt[:LEVel]? [MIN|MAX,] (@<chanlist>)
Specifies the initial current, voltage, power, or resistance before the staircase occurs.
[SOURce:]ARB:SEQuence:STEP:CURRent:STAircase:STARt[:LEVel] <value>, <step#>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:CURRent:STAircase:STARt[:LEVel]? [MIN|MAX,] <step#>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:VOLTage:STAircase:STARt[:LEVel] <value>, <step#>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:VOLTage:STAircase:STARt[:LEVel]? [MIN|MAX,] <step#>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:POWer:STAircase:STARt[:LEVel] <value>, <step#>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:POWer:STAircase:STARt[:LEVel]? [MIN|MAX,] <step#>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:RESISTANCE:STAircase:STARt[:LEVel] <value>, <step#>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:RESISTANCE:STAircase:STARt[:LEVel]? [MIN|MAX,] <step#>, (@<chanlist>)
Specifies the initial current, voltage, power, or resistance before the staircase occurs for Arb sequence
steps.
The difference between the start level and the end level is divided equally among the
number of steps.
5 SCPI Programming Reference
208 Keysight N6700C Operating and Service Guide