List Commands
List commands program an output sequence of multiple current, voltage, power, or resistance
settings. A comma-delimited list of up to 512 steps may be programmed. Note that for Models N678xA
SMU , only the parameters associated with one of the priority modes (either voltage
priority or current priority) may be list controlled.
LIST commands require to be installed on models N673xA. N674xA,
N675xA, and N677xA.
[SOURce:]LIST:COUNt <value>, (@<chanlist>)
[SOURce:]LIST:COUNt? [MIN|MAX,] (@<chanlist>)
Sets the list repeat count. This sets the number of times that a list is executed before it completes. The
count range is 1 through 4096. Use the Infinity parameter to run the list continuously.
Parameter Typical Return
1 ā 4096|MIN|MAX|INFinity, *RST 1 <list count>
Sets the list count to 10: LIST:COUN 10, (@1)
l Use ABORt:TRANsient to stop the list at any time. When the list is aborted, the output returns to
the settings that were in effect before the list started.
l The query returns the list count for each channel specified. If a repeat count of 9.9E37 is returned,
it means the list is set to repeat continuously.
[SOURce:]LIST:CURRent[:LEVel] <value>{,<value>}, (@<chanlist>)
[SOURce:]LIST:CURRent[:LEVel]? (@<chanlist>)
[SOURce:]LIST:VOLTage[:LEVel] <value>,{<value>}, (@<chanlist>)
[SOURce:]LIST:VOLTage[:LEVel]? (@<chanlist>)
[SOURce:]LIST:POWer[:LEVel] <value>,{<value>}, (@<chanlist>)
[SOURce:]LIST:POWer[:LEVel]? (@<chanlist>)
[SOURce:]LIST:RESistance[:LEVel] <value>,{<value>}, (@<chanlist>)
[SOURce:]LIST:RESistance[:LEVel]? (@<chanlist>)
Specifies the setting for each list step.ī Values are specified in amperes, volts, watts, or ohms. A
comma-delimited list of up to 512 steps may be programmed.
Parameter Typical Return
0 to maximum. Values are range and module-rating dependent.
*RST 1 step set to the minimum programmable value.
<list value 1>,
<list value 2>,
<list value 3>...
5 SCPI Programming Reference
256 Keysight N6700C Operating and Service Guide