l With continuous triggering disabled, the trigger system must be initiated for each trigger using the
INITiate:TRANsient command.
l ABORt:TRANsient does not abort continuous triggers if INITiate:CONTinuous:TRANsient ON has
been programmed. In this case, the trigger system will automatically re-initiate.
Keysight N6700C Operating and Service Guide 255