Function Command
[SOURce:]FUNCtion CURRent|VOLTage|POWer|RESistance, (@<chanlist>)
[SOURce:]FUNCtion? (@<chanlist>)
Sets the output priority mode). Power and Resistance priority modes only apply to models N679xA.
In voltage priority mode the output is controlled by a bi-polar constant voltage feedback loop, which
maintains the output voltage at its positive or negative programmed setting. In current priority mode
the output is controlled by a bi-polar constant current feedback loop, which maintains the output
sourcing or sinking current at its programmed setting.
For N679xA load modules in power or resistance priority modes, the constant current along with the
constant voltage feedback loops maintain the input power and resistance at their programmed
settings.
Refer to Operating Mode Tutorial for more information.
Parameter Typical Return
CURRent|VOLTage|POWer|RESistance,
*RST CURRent for N679xA
*RST VOLTage for N678xA SMU
CURR, VOLT, POW, or RES
Selects current priority regulation mode: FUNC CURR, (@1)