EasyManua.ls Logo

Keysight N6702C - Trigger Commands

Keysight N6702C
430 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Trigger Commands
Trigger commands control the transient and acquisition subsystems.
All TRIGger commands except for TRIGger:TRANsient[:IMMediate] and
TRIGger:TRANsient:SOURce require to be installed on models N673xA.
N674xA, N675xA, and N677xA.
TRIGger:ACQuire[:IMMediate] (@<chanlist>)
TRIGger:ELOG[:IMMediate] (@<chanlist>)
TRIGger:TRANsient[:IMMediate] (@<chanlist>)
These commands override any selected trigger source and generate immediate triggers.
TRIGger:ACQuire - Generates an immediate measurement trigger.
TRIGger:ELOG - Generates an immediate trigger for external data logging.
TRIGger:TRANsient - Generates an immediate output transient trigger.
Parameter Typical Return
(none) (none)
Generates an acquisition trigger: TRIG:ACQ (@1)
l You must initiate the applicable trigger system before you can send any trigger.
l At acquire trigger completion, the WTG-meas bit in the Status Operation Condition register is
cleared.
l At transient trigger completion, the WTG-tran bit in the Status Operation Condition register is
cleared.
TRIGger:ACQuire:CURRent[:LEVel] <value>, (@<chanlist>)
TRIGger:ACQuire:CURRent[:LEVel]? [MIN|MAX,] (@<chanlist>)
TRIGger:ACQuire:VOLTage[:LEVel] <value>, (@<chanlist>)
TRIGger:ACQuire:VOLTage[:LEVel]? [MIN|MAX,] (@<chanlist>)
Sets the triggered level of the output. Applies when the measurement trigger source is set to a level.
Values are specified in either amperes or volts. The MIN and MAX values depend on the ratings of
the unit.
Parameter Typical Return
Current: 0 - maximum|MIN|MAX, *RST 0
Voltage: 0 - maximum|MIN|MAX, *RST 0
The maximum values are 102% of the rating of the module.
<triggered level>
Set the triggered current level to 3 A: TRIG:ACQ:CURR 3, (@1)
Set the triggered voltage level to 50 V: TRIG:ACQ:VOLT50, (@1)
Keysight N6700C Operating and Service Guide 297
5 SCPI Programming Reference

Table of Contents

Related product manuals