[SOURce:]ARB:CURRent:TRAPezoid:FTIMe <value>, (@<chanlist>)
[SOURce:]ARB:CURRent:TRAPezoid:FTIMe? [MIN|MAX,] (@<chanlist>)
[SOURce:]ARB:VOLTage:TRAPezoid:FTIMe<value>, (@<chanlist>)
[SOURce:]ARB:VOLTage:TRAPezoid:FTIMe? [MIN|MAX,] (@<chanlist>)
[SOURce:]ARB:POWer:TRAPezoid:FTIMe<value>, (@<chanlist>)
[SOURce:]ARB:POWer:TRAPezoid:FTIMe? [MIN|MAX,] (@<chanlist>)
[SOURce:]ARB:RESISTANCE:TRAPezoid:FTIMe<value>, (@<chanlist>)
[SOURce:]ARB:RESISTANCE:TRAPezoid:FTIMe? [MIN|MAX,] (@<chanlist>)
Specifies the fall time of the trapezoid in seconds.
[SOURce:]ARB:SEQuence:STEP:CURRent:TRAPezoid:FTIMe <value>, <step#>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:CURRent:TRAPezoid:FTIMe? [MIN|MAX,] <step#>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:VOLTage:TRAPezoid:FTIMe <value>, <step#>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:VOLTage:TRAPezoid:FTIMe? [MIN|MAX,] <step#>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:POWer:TRAPezoid:FTIMe <value>, <step#>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:POWer:TRAPezoid:FTIMe? [MIN|MAX,] <step#>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:RESISTANCE:TRAPezoid:FTIMe <value>, <step#>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:RESISTANCE:TRAPezoid:FTIMe? [MIN|MAX,] <step#>, (@<chanlist>)
Specifies the fall time of the trapezoid in seconds for Arb sequence steps.
Parameter Typical Return
0 – 262.144 | MIN | MAX, *RST 1 <fall time>
Step 0 - 99, *RST 0 (none)
Programs a fall time for the current trapezoid: ARB:CURR:TRAP:FTIM 10,(@1)
Programs a fall time for the voltage trapezoid: ARB:VOLT:TRAP:FTIM 10,(@1)
Programs a fall time for the power trapezoid: ARB:POW:TRAP:FTIM 10,(@1)
Programs a fall time for the resistance trapezoid: ARB:RESISTANCE:TRAP:FTIM 10,(@1)
Programs a fall time for step 0:ARB:SEQ:STEP:CURR:TRAP:FTIM 10, 0,(@1)
[SOURce:]ARB:CURRent:TRAPezoid:RTIMe <value>, (@<chanlist>)
[SOURce:]ARB:CURRent:TRAPezoid:RTIMe? [MIN|MAX,] (@<chanlist>)
[SOURce:]ARB:VOLTage:TRAPezoid:RTIMe<value>, (@<chanlist>)
[SOURce:]ARB:VOLTage:TRAPezoid:RTIMe? [MIN|MAX,] (@<chanlist>)
[SOURce:]ARB:POWer:TRAPezoid:RTIMe<value>, (@<chanlist>)
[SOURce:]ARB:POWer:TRAPezoid:RTIMe? [MIN|MAX,] (@<chanlist>)
[SOURce:]ARB:RESISTANCE:TRAPezoid:RTIMe<value>, (@<chanlist>)
[SOURce:]ARB:RESISTANCE:TRAPezoid:RTIMe? [MIN|MAX,] (@<chanlist>)
Specifies the rise time of the trapezoid in seconds.
Keysight N6700C Operating and Service Guide 215
5 SCPI Programming Reference