[SOURce:]ARB:CURRent:CDWell:DWELl <value>,(@<chanlist>)
[SOURce:]ARB:CURRent:CDWell:DWELl? (@<chanlist>)
[SOURce:]ARB:VOLTage:CDWell:DWELl <value>,(@<chanlist>)
[SOURce:]ARB:VOLTage:CDWell:DWELl? (@<chanlist>)
[SOURce:]ARB:POWer:CDWell:DWELl <value>,(@<chanlist>)
[SOURce:]ARB:POWer:CDWell:DWELl? (@<chanlist>)
[SOURce:]ARB:RESISTANCE:CDWell:DWELl <value>,(@<chanlist>)
[SOURce:]ARB:RESISTANCE:CDWell:DWELl? (@<chanlist>)
Specifies the dwell time in seconds for each current, voltage, power, or resistance point in a constant-
dwell Arb
You can only program one constant-dwell Arb type at a time, so setting the current CD Arb resets the
voltage, power, and resistance Arbs to 1 point at the minimum value. Programmed values can range
from 10.24 microseconds to 0.30 seconds. Values are rounded to the nearest 10.24-microsecond
increment.
Parameter Typical Return
0.00001024 - 0.30, *RST 0.001 <dwell value>
Program a constant-dwell current time: ARB:CURR:CDW:DWEL 0.2,(@1)
Program a constant-dwell voltage time: ARB:VOLT:CDW:DWEL 0.5,(@1)
Program a constant-dwell power time: ARB:POW:CDW:DWEL 5,(@1)
Program a constant-dwell resistance time: ARB:RESISTANCE:CDW:DWEL 2,(@1)
l The maximum repeat count for CD Arbs is 256. See [SOURce:]ARB:COUNt.
[SOURce:]ARB:CURRent:CDWell:POINts? (@<chanlist>)
[SOURce:]ARB:VOLTage:CDWell:POINts? (@<chanlist>)
[SOURce:]ARB:POWer:CDWell:POINts? (@<chanlist>)
[SOURce:]ARB:RESISTANCE:CDWell:POINts? (@<chanlist>)
Returns the number of current, voltage, power, or resistance points in a constant-dwell Arb in the form
+n.nnnnnnE+nn.
Parameter Typical Return
(none) <number of points>
Program a constant-dwell time of 0.2 seconds:
ARB:CURR:CDW:DWEL 0.2,(@1)
ARB:VOLT:CDW:DWEL 0.2,(@1)
ARB:POW:CDW:DWEL 0.2,(@1)
ARB:RESISTANCE:CDW:DWEL 0.2,(@1)
Keysight N6700C Operating and Service Guide 189
5 SCPI Programming Reference