[SOURce:]ARB:CURRent:PULSe:FREQuency<value>, (@<chanlist>)
[SOURce:]ARB:CURRent:PULSe:FREQuency? [MIN|MAX,] (@<chanlist>)
[SOURce:]ARB:VOLTage:PULSe:FREQuency <value>, (@<chanlist>)
[SOURce:]ARB:VOLTage:PULSe:FREQuency? [MIN|MAX,] (@<chanlist>)
[SOURce:]ARB:POWer:PULSe:FREQuency <value>, (@<chanlist>)
[SOURce:]ARB:POWer:PULSe:FREQuency? [MIN|MAX,] (@<chanlist>)
[SOURce:]ARB:RESISTANCE:PULSe:FREQuency <value>, (@<chanlist>)
[SOURce:]ARB:RESISTANCE:PULSe:FREQuency? [MIN|MAX,] (@<chanlist>)
Specifies the frequency of the current, voltage, power, or resistance pulse in Hertz. The frequency
commands are coupled with the pulse time commands.
[SOURce:]ARB:SEQuence:STEP:CURRent:PULSe:FREQuency<value>, <step#>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:CURRent:PULSe:FREQuency? [MIN|MAX,] <step#>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:VOLTage:PULSe:FREQuency <value>, <step#>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:VOLTage:PULSe:FREQuency? [MIN|MAX,] <step#>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:POWer:PULSe:FREQuency <value>, <step#>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:POWer:PULSe:FREQuency? [MIN|MAX,] <step#>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:RESISTANCE:PULSe:FREQuency <value>, <step#>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:RESISTANCE:PULSe:FREQuency? [MIN|MAX,] <step#>, (@<chanlist>)
Specifies the frequency of the current, voltage, power, or resistance pulse in Hertz for Arb sequence
steps. The frequency commands are coupled with the pulse time commands.
Parameter Typical Return
3.8147E-5 - 1.0E4 | MIN | MAX, *RST 1 <frequency>
Step 0 - 99, *RST 0 (none)
Programs a current frequency: ARB:CURR:PULS:FREQ 10,(@1)
Programs a voltage frequency: ARB:VOLT:PULS:FREQ 10,(@1)
Programs a power frequency: ARB:POW:PULS:FREQ 10,(@1)
Programs a resistance frequency: ARB:RESISTANCE:PULS:FREQ 10,(@1)
Programs a current frequency for step 0: ARB:SEQ:STEP:CURR:PULS:FREQ 10, 0,(@1)
[SOURce:]ARB:CURRent:PULSe:STARt[:LEVel] <value>, (@<chanlist>)
[SOURce:]ARB:CURRent:PULSe:STARt[:LEVel]? [MIN|MAX,] (@<chanlist>)
[SOURce:]ARB:VOLTage:PULSe:STARt[:LEVel] <value>, (@<chanlist>)
[SOURce:]ARB:VOLTage:PULSe:STARt[:LEVel]? [MIN|MAX,] (@<chanlist>)
[SOURce:]ARB:POWer:PULSe:STARt[:LEVel] <value>, (@<chanlist>)
[SOURce:]ARB:POWer:PULSe:STARt[:LEVel]? [MIN|MAX,] (@<chanlist>)
[SOURce:]ARB:RESISTANCE:PULSe:STARt[:LEVel] <value>, (@<chanlist>)
[SOURce:]ARB:RESISTANCE:PULSe:STARt[:LEVel]? [MIN|MAX,] (@<chanlist>)
Specifies the current, voltage, power, or resistance before and after the pulse occurs.