EasyManua.ls Logo

Keysight N6702C - Page 204

Keysight N6702C
430 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
[SOURce:]ARB:CURRent:SINusoid:FREQuency<value>, (@<chanlist>)
[SOURce:]ARB:CURRent:SINusoid:FREQuency? [MIN|MAX,] (@<chanlist>)
[SOURce:]ARB:VOLTage:SINusoid:FREQuency <value>, (@<chanlist>)
[SOURce:]ARB:VOLTage:SINusoid:FREQuency? [MIN|MAX,] (@<chanlist>)
[SOURce:]ARB:POWer:SINusoid:FREQuency <value>, (@<chanlist>)
[SOURce:]ARB:POWer:SINusoid:FREQuency? [MIN|MAX,] (@<chanlist>)
[SOURce:]ARB:RESISTANCE:SINusoid:FREQuency <value>, (@<chanlist>)
[SOURce:]ARB:RESISTANCE:SINusoid:FREQuency? [MIN|MAX,] (@<chanlist>)
Specifies the frequency of the current, voltage, power, or resistance sine wave in Hertz.
[SOURce:]ARB:SEQuence:STEP:CURRent:SINusoid:FREQuency<value>, <step#>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:CURRent:SINusoid:FREQuency? [MIN|MAX,] <step#>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:VOLTage:SINusoid:FREQuency <value>, <step#>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:VOLTage:SINusoid:FREQuency? [MIN|MAX,] <step#>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:POWer:SINusoid:FREQuency <value>, <step#>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:POWer:SINusoid:FREQuency? [MIN|MAX,] <step#>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:RESISTANCE:SINusoid:FREQuency <value>, <step#>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:RESISTANCE:SINusoid:FREQuency? [MIN|MAX,] <step#>, (@<chanlist>)
Specifies the frequency of the current, voltage, power, or resistance sine wave in Hertz for Arb
sequence steps.
Parameter Typical Return
3.8147E-5 - 1.0E4 | MIN | MAX, *RST 1 <frequency>
Step 0 - 99, *RST 0 (none)
Programs a current frequency: ARB:CURR:SIN:FREQ 10,(@1)
Programs a voltage frequency: ARB:VOLT:SIN:FREQ 10,(@1)
Programs a power frequency: ARB:POW:SIN:FREQ 10,(@1)
Programs a resistance frequency: ARB:RESISTANCE:SIN:FREQ 10,(@1)
Programs a current frequency for step 0: ARB:SEQ:STEP:CURR:SIN:FREQ 10, 0,(@1)
[SOURce:]ARB:CURRent:SINusoid:OFFSet<value>, (@<chanlist>)
[SOURce:]ARB:CURRent:SINusoid:OFFSet? [MIN|MAX,] (@<chanlist>)
[SOURce:]ARB:VOLTage:SINusoid:OFFSet <value>, (@<chanlist>)
[SOURce:]ARB:VOLTage:SINusoid:OFFSet? [MIN|MAX,] (@<chanlist>)
[SOURce:]ARB:POWer:SINusoid:OFFSet <value>, (@<chanlist>)
[SOURce:]ARB:POWer:SINusoid:OFFSet? [MIN|MAX,] (@<chanlist>)
[SOURce:]ARB:RESISTANCE:SINusoid:OFFSet <value>, (@<chanlist>)
[SOURce:]ARB:RESISTANCE:SINusoid:OFFSet? [MIN|MAX,] (@<chanlist>)
Specifies the offset of the current, voltage, power, or resistance sine wave from zero.
5 SCPI Programming Reference
204 Keysight N6700C Operating and Service Guide

Table of Contents

Related product manuals