Parameter Typical Return
(none) <HIGH value>
Returns the measured high level current FETC:CURR:HIGH? (@1)
Returns the measured high level voltage FETC:VOLT:HIGH? (@1)
FETCh[:SCALar]:CURRent:LOW? (@<chanlist>)
FETCh[:SCALar]:VOLTage:LOW? (@<chanlist>)
Returns the Low level of a pulse waveform. Values returned are either in amperes, or volts. See
Measurement Types. The reading is in the form +n.nnnnnnE+nn for each channel specified. Multiple
responses are separated by commas.
Parameter Typical Return
(none) <LOWvalue>
Returns the measured low level current FETC:CURR:LOW? (@1)
Returns the measured low level voltage FETC:VOLT:LOW? (@1)
FETCh[:SCALar]:CURRent:MAXimum? (@<chanlist>)
FETCh[:SCALar]:VOLTage:MAXimum? (@<chanlist>)
FETCh[:SCALar]:POWer:MAXimum? (@<chanlist>)
FETCh[:SCALar]:CURRent:MINimum? (@<chanlist>)
FETCh[:SCALar]:VOLTage:MINimum? (@<chanlist>)
FETCh[:SCALar]:POWer:MINimum? (@<chanlist>)
Returns the maximum or minimum value. Values returned are either in amperes, volts, or Watts. See
Measurement Types. The reading is in the form +n.nnnnnnE+nn for each channel specified. Multiple
responses are separated by commas.
Parameter Typical Return
(none) <MIN value>
<MAX value>
Returns the measured minimum current FETC:CURR:MIN? (@1)
Returns the measured minimum voltage FETC:VOLT:MIN? (@1)
Returns the measured minimum power FETC:POW:MIN? (@1)
Returns the measured maximum current FETC:CURR:MAX? (@1)
Returns the measured maximum voltage FETC:VOLT:MAX? (@1)
Returns the measured maximum power FETC:POW:MAX? (@1)
FETCh:ARRay:CURRent[:DC]? (@<chanlist>)
FETCh:ARRay:VOLTage[:DC]? (@<chanlist>)
FETCh:ARRay:POWer[:DC]? (@<chanlist>)
Returns an array with the instantaneous measurement.Values returned are either in amperes, volts,
or watts.
Keysight N6700C Operating and Service Guide 243
5 SCPI Programming Reference