l The query command returns 0 if the output returns to the settings that were in effect before the list
started, and 1 if the output voltage or current remains at the value of the last list step.
[SOURce:]LIST:TOUTput:BOSTep[:DATA] 0|OFF|1|ON {,0|OFF|1|ON}, (@<chanlist>)
[SOURce:]LIST:TOUTput:BOSTep[:DATA]? (@<chanlist>)
[SOURce:]LIST:TOUTput:EOSTep[:DATA] 0|OFF|1|ON {,0|OFF|1|ON}, (@<chanlist>)
[SOURce:]LIST:TOUTput:EOSTep[:DATA]? (@<chanlist>)
Specifies which list steps generate a trigger signal at the beginning of step (BOSTep) or end of step
(EOSTep). A comma-delimited list of up to 512 steps may be programmed.
A trigger is only generated when the state is set to ON. This signal can be used as a trigger source for
measurements and transients of other units, and for digital pins configured as trigger outputs.
Parameter Typical Return
0|OFF|1|ON ,
*RST 1 step with a value of OFF
<value 1>,<value 2>,<value 3>...
To generatetriggers at the beginning of the second step of a 3-step list:
LIST:TOUT:BOST OFF,ON,OFF, (@1)
l The order in which the list values are entered determines the sequence when the list executes.
l To create a valid list, the current, voltage, power, resistance, BOST, EOST, and dwell lists must all
be the same length or have a length of 1, which is interpreted as having the same length as the
longest list.
l This command overwrites any present list in active memory; it does not append to the present list.
Keysight N6700C Operating and Service Guide 259