Parameter Typical Return
To resetthe front panel password: SYST:PASS:FPAN:RES
SYSTem:PERSona:MANufacturer "Agilent"|"Keysight"
SYSTem:PERSona:MANufacturer?
SYSTem:PERSona:MANufacturer:DEFault
Changes the manufacturer identity. This command is used to change the manufacturer's ID string for
backward compatibility. This affects the following identity items:
l *IDN? contains "Agilent" instead of "Keysight"
l Default LAN hostname has the prefix A instead of K
l The LXI XML file fields include "Agilent" instead of "Keysight"
l The mDNS service name contains "Agilent" instead of "Keysight"
The DEFault command returns the manufacturer's ID back to Keysight.
Parameter Typical Return
"Agilent"|"Keysight" Keysight
Changes the manufacturer's ID to Agilent: SYST:PERS:MAN "Agilent"
SYSTem:PERSona:MODel "<model number>"
SYSTem:PERSona:MODel?
SYSTem:PERSona:MODel:DEFault
Changes the model number. This command is used to change the mainframe model ID string for
backward compatibility.
Compatibility Factory setting
N6700B
N6701A
N6702A
N6700C
N6701C
N6702C
The DEFault command returns the model's ID to the factory setting.
Parameter Typical Return
"<model number>" N6700C, N6701C, or N6702C
Changes the model ID to N6700B: SYST:PERS:MOD "N6700B"
SYSTem:REBoot
Reboots the instrument to its power-on state. Refer to Reset State for more information.
Keysight N6700C Operating and Service Guide 295
5 SCPI Programming Reference