A triggered value is on a different range than the one that is presently set.
306 Too many list points
Too many list points have been specified.
307 List lengths are not equivalent
One or more lists are not the same length.
308 This setting cannot be changed while transient trigger is initiated
Setting cannot be changed while the instrument is waiting for or executing a trigger sequence.
309 Cannot initiate, voltage and current in fixed mode
Cannot initiate transient generator. Either the voltage or current function is set to Fixed mode.
310 The command is not supported by this model
This power modules either does not have the hardware capability or the options required to support
this command.
311 This group of channel has non-identical features
One or more modules in this group does not match the ratings or options of the other modules.
312 The command is disabled when the channel is grouped
This command is disabled when the output channel is grouped.
315 Settings conflict error
A data element could not be programmed because of the present instrument state.
320 Firmware update error
This may be due to the instrument hardware not being able to support the firmware version.
Command Errors (these errors set Standard Event Status register bit #5)
-100 Command error
Generic syntax error.
-101 Invalid character
An invalid character was found in the command string.
-102 Syntax error
Invalid syntax was found in the command string. Check for blank spaces.
-103 Invalid separator
An invalid separator was found in the command string. Check for proper use of , ; :
-104 Data type error