Low Range Current Programming - Models N6761A through N6766A
Step 1. Connect only the current measurement terminals of the Keysight 3458A to the output
terminals (see Cal Setup D without the resistor).
Step 2. Select the low-current programming range. The value to program a range must be the
maximum current of the range. This example selects the 100 mA range of output 1.
Front Panel Menu Reference SCPI Command
Select System\Admin\Cal\Curr\Iprog
Select the Low range. Then press Next.
CAL:CURR 0.1, (@1)
Step 3. Select the first current calibration point and enter the data. For models N6761A and N6762A
only, wait a minimum of 5 minutes for the output current to stabilize. before entering the data.
Front Panel Menu Reference SCPI Command
Display shows: “Enter P1 measured data” Measure the current with the Keysight
3458A and enter the data.
Press Select when done.
CAL:LEV P1
*OPC?
CAL:DATA <data>
Step 4. Select the second current calibration point and enter the data. Wait a minimum of 30 seconds
for the output current to stabilize
Front Panel Menu Reference SCPI Command
Display shows: “Enter P2 measured data”
Measure the current with the Keysight 3458A and enter the data. Press Select
when done. Press Next to finish calibration.
CAL:LEV P2
*OPC?
CAL:DATA <data>
Low Range Current Measurement - Models N6761A through N6766A
Step 1. Connect only the current measurement terminals of the Keysight 3458A to the output
terminals (see Cal Setup D without the resistor).
Step 2. Select the low-current measurement range. The value to program a range must be the
maximum current of the range. This example selects the 100 mA range of output 1.
Front Panel Menu Reference SCPI Command
Select System\Admin\Cal\Curr\Imeas
Select the Low1 range. Then press Next.
CAL:CURR:MEAS 0.1, (@1)
Step 3. Select the first current calibration point and enter the data. For models N6761A and N6762A
only, wait a minimum of 5 minutes for the output current to stabilize before entering the data.
Front Panel Menu Reference SCPI Command
Display shows: “Enter measured data”
Measure the current with the Keysight 3458A and enter the data. Press Select
when done.
CAL:LEV P1
*OPC?
CAL:DATA <data>