Front Panel Menu Reference SCPI Command
Display shows: “Enter measured data”
Measure the current with the Keysight 3458A and enter the data. Press Select
when done.
CAL:LEV P1
*OPC?
CAL:DATA <data>
Negative Current Limit Calibration
Step 1. Connect the Keysight 3458A, current shunt, and an external power supply to the output (see
Cal Setup H). Connect plus to plus, minus to minus. Set the external power supply to 3V, 3A.
Step 2. Select the negative-current range. The value to select the range must be negative and the
maximum current of the range.
Front Panel Menu Reference SCPI Command
Select System\Admin\Cal\Curr\Ilim CAL:CURR:LIM:NEG -2, (@1)
Step 3. Select the first current calibration point and enter the data. This value should be negative.
Front Panel Menu Reference SCPI Command
Display shows: “Enter P1 measured data” Calculate the shunt current (I=V/R)
and enter the data.
Press Select when done.
CAL:LEV P1
*OPC?
CAL:DATA <data>
Step 4. Select the second current calibration point and enter the data. This value should be negative.
Front Panel Menu Reference SCPI Command
Display shows: “Enter P2 measured data”
Calculate the shunt current (I=V/R) and enter the data. Press Selectwhen done.
Press Next to finish calibration.
CAL:LEV P2
*OPC?
CAL:DATA <data>
Enter a Calibration Date, Save, and Log Out
Front Panel Menu Reference SCPI Command
Select System\Admin\CAL\DATE.
Enter the calibration date in the Datefield.
Select System\Admin\CAL\Save.
Select Save to save all calibration data.
Select System\Admin\Logout
to exitcalibration mode.
To enter a calibration date:
CAL:DATE "<date>"
To save the calibration data:
CAL:SAVE
To exit calibration mode:
CAL:STAT OFF
6 Verification and Calibration
372 Keysight N6700C Operating and Service Guide