EasyManua.ls Logo

Keysight N9010B - Page 510

Keysight N9010B
2852 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
3 Spectrum Analyzer Mode
3.2 Swept SA Measurement
On the top line each trace number is shown, in the trace color. A blue box is drawn
around the currently selected trace.
Below each trace number, is a letter signifying the trace type for that trace number,
where
W Clear/Write
A Trace Average
M Max Hold
m Min Hold
If the letter is white, it means the trace is being updated (Update = On); if the letter is
dimmed , it means the trace is not being updated (Update = Off). A strikethrough (for
example, W) indicates that the trace is blanked (Display = Off). Note that it is
possible for a trace to be updating and blanked, which is useful if the trace is a trace
math component.
The third line shows the detector type for each trace, or, if trace math is on for that
trace, it shows f (for “Math Function”). It is not always possible to have a unique
detector for each trace, but the instrument hardware provides the maximum
flexibility of detector selection to maintain the highest accuracy.
The letters used for this readout are
N Normal
A Average
P Peak
p Negative Peak
S Sample
Q Quasi Peak
E EMI Average
R RMS Average
f Math Function
If the DET letter is green, it means the detector is in Auto; if it is white, it means the
detector has been manually selected.
So, in the example above, the panel shows the following:
Trace 1 Visible, being updated, in Clear/Write, with Normal detector auto selected
Trace 2 Visible, being updated, in Clear/Write, being written to with a math function
Trace 3 Visible, not updating, data was taken in Max Hold, with Peak detector auto selected
510 Spectrum Analyzer Mode User's &Programmer's Reference

Table of Contents

Other manuals for Keysight N9010B

Related product manuals