137
Command Reference
Reflection Measurement Option Subsystem
Reflection Measurement Center Frequency
[:SENSe]:CAT:FREQuency:CENTer <freq>
[:SENSe]:CAT:FREQuency:CENTer?
Use this command to define the center frequency of
reflection measurement.
Range: 5 MHz to 7 GHz
Example: :CAT:FREQuency:CENTer 200 MHz
Key access: Freq > Center Freq
Reflection Measurement Frequency Span
[:SENSe]:CAT:FREQuency:SPAN <freq>
[:SENSe]:CAT:FREQuency:SPAN?
This command defines the frequency span of current
measurement.
Example: ;CAT:FREQuency:SPAN 200 MHz
Key access: Span > Span
Reflection Measurement Distance Start
[:SENSe]:CAT:DISTance:STARt <distance>
[:SENSe]:CAT:DISTance:STARt?
This command defines the start distance point of DTF
measurement
Range: 0 to Stop distance - 0.1
Example: :CAT:DISTance:STARt 1.3
Key access: Meas > Start Distance
Reflection Measurement Distance Stop
[:SENSe]:CAT:DISTance:STOP <distance>
[:SENSe]:CAT:DISTance:STOP?
This command defines the stop distance of DTF
measurement
Range: 0 to stop distance – 0.1 meter
Key access: Meas > Stop Distance