EasyManua.ls Logo

Keysight N9322C - Page 186

Keysight N9322C
223 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Command Reference
Demodulation Option Subsystem
178
FM X Scale/DIV
:DISPlay:AMA:VIEW:WINDow:TRACe:X[:SCALe]:PDIVi-
sion <real_time>
:DISPlay:AMA:VIEW:WINDow:TRACe:X[:SCALe]:PDIVi-
sion?
This command specifies scale per division of X axis.
Range: 1e–1 to 1
Example: :DISPlay:AMA:VIEW:WINDow:TRACe:X:PDIVision 1e8
Key access: Shift > Disp > X Scale > Scale/DIV
FM X Reference Value
:DISPlay:AMA:VIEW:WINDow:TRACe:X[:SCALe]:RVALue
< real_time >
:DISPlay:AMA:VIEW:WINDow:TRACe:X[:SCALe
]:RVALue?
This command specifies the reference value of X axis.
Range: –5 to +5
Example: :DISPlay:AMA:VIEW:WINDow:TRACe:X:RVALue 1e8
Key access: Shift > Disp > X Scale > Ref Value
FM X Reference Position
:DISPlay:AMA:VIEW:WINDow:TRACe:X[:SCALe]:RPOSiti
on LEFT|CENTer|RIGHt
:DISPlay:AMA:VIEW:WINDow:TRACe:X[:SCALe]:RPOSiti
on?
This command specifies the reference position of X axis.
Example: :DISPlay:AMA:VIEW:WINDow:TRACe:X:RPOSition LEFT
Key access: Shift > Disp > X Scale > Ref Position

Table of Contents