Chapter 2 29
Test Descriptions
Input VSWR
Measuring VSWR from 10 MHz to 500 MHz
Step 1. Connect the test equipment as shown in Figure 2-4 on page 28.
Step 2. Set the Network Analyzer Start/Stop frequencies from 10 MHz to
500 MHz.
Press the
Start key then 1, 0, M/µ; then press the Stop key and 5, 0, 0 M/µ.
Step 3. On the NFA ensure that the preset is set to the factory settings (press
the
System key and More, Power On/Preset, Power On (Preset) and
Preset (Factory)).
Step 4. On the NFA press the
Preset key and wait for the routine to finish
Step 5. Set sweep to single (press the
Sweep key and Sweep Mode: Single).
Step 6. Set the frequency mode to fixed (press the
Frequency/Points key and
Freq Mode, Fixed).
Step 7. Set the fixed frequency to 250 MHz (press the
Frequency/Points key and
Fixed Freq, 2, 5, 0, MHz).
Step 8. Restart the measurement sweep on the DUT (press the
Restart key).
Step 9. Set the Network Analyzer to autorange (press the
Scale Ref key and
Auto Scale).
Step 10. Using the marker peak function on the network analyzer, determine the
maximum displayed VSWR over the measured frequency range (press
the
Marker Fctn key and Mkr Search [ON], Search: Max).
Step 11. Record the maximum displayed signal frequency and VSWR on the Test
Record.
Step 12. Verify that the measured VSWR from 10 MHz to 500 MHz is within the
published specification.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight NFA Series and is the answer not in the manual?

Keysight NFA Series Specifications

General IconGeneral
BrandKeysight
ModelNFA Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals