EasyManuals Logo

Keysight NFA Series User Manual

Keysight NFA Series
201 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #54 background imageLoading...
Page #54 background image
34 Chapter 2
Test Descriptions
Input VSWR
Figure 2-5 Vector Network Analyzer: Adaptor Connection
Step 9. As the calibration kit selection is dependant on the NFA being tested,
select the relevant calibration kit in use from the Cal menu (press the
Cal
key,
Calibration Kit selection and Return).
NOTE The connector type applies to the test port or adaptor if fitted and not the
NFA under test.
Step 10. Select an S11 port calibration (press the
Cal key, then Calibrate Menu and
Reflection 1-port).
Step 11. Connect the Open to the test port or, if fitted, the adapter and perform
the open calibration (press
Opens and then select the relevant menu
key).
Step 12. Connect the Short to the test port or, if fitted, the adapter and perform
the open calibration (press
Shorts and then select the relevant menu
key).
Step 13. Connect the Load to the test port or, if fitted, the adapter and perform
the open calibration (press
Load, select the relevant menu key and then
done 1-port CAL
).
Step 14. Save the calibration (press the
Save/recall key and then Save State).
VECTOR NETWORK
ANALYZER 2
Port 1
Adapter 3
uWave Cable

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight NFA Series and is the answer not in the manual?

Keysight NFA Series Specifications

General IconGeneral
BrandKeysight
ModelNFA Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals