Chapter 2 35
Test Descriptions
Input VSWR
Measuring Input VSWR from 3000 MHz to 6700 MHz
Figure 2-6 Input VSWR Test Setup 2
Step 1. Connect the test equipment as shown in Figure 2-6 on page 35.
Step 2. Set the Network Analyzer Start/Stop to measure the DUT’s input VSWR
from 3000 MHz to 6700 MHz.
Press the
Start key and then 3,., 0, G/n; then press the Stop key and 6, ., 7,
G/n
Step 3. Set the fixed frequency of the NFA to 5000 MHz (press the
Frequency/Points key and then Fixed Freq, 5, GHz).
Step 4. Restart the measurement sweep on the DUT (press the
Restart key).
Step 5. Using the marker peak function on the network analyzer, determine the
maximum displayed VSWR over the measured frequency range (press
the
Marker Fctn key, Mkr Search [ON] then Search: Max).
Step 6. Record the maximum displayed signal frequency and VSWR on the Test
Record.
Step 7. Verify that the measured VSWR from 3000 MHz to 6700 MHz is within
the published specification.
VECTOR NETWORK
ANALYZER 2
NOISE FIGURE
ANALYZER
Port 1
Adapter 3
uWave Cable
Input