EasyManua.ls Logo

Keysight U1271A - AC+DC Specifications for U1272 A; Table 5-6 U1272 A True Rms AC+DC Voltage Specifications

Keysight U1271A
188 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Characteristics and Specifications 5
Electrical Specifications
U1271A/U1272A User’s Guide 165
AC+DC specifications for U1272A
Table 5-6 U1272A true rms AC+DC voltage specifications
Function Range Resolution
Accuracy
45 Hz to
65 Hz
20 Hz to
45 Hz
1 kHz to
5 kHz
5 kHz to
20 kHz
20 kHz to
100 kHz
[3]
65 Hz to
1 kHz
Voltage
30 mV 0.001 mV 0.7% + 40 0.8% + 45 1.1% + 45 1.1% + 60 3.6% + 60
300 mV 0.01 mV 0.7% + 25 0.8% + 30 1.1% + 30 1.1% + 45 3.6% + 45
3 V 0.0001 V 0.7% + 25 1.1% + 30 1.6% + 30 2.1% + 45 3.6% + 45
30 V 0.001 V 0.7% + 25 1.1% + 30 1.6% + 30 2.1% + 45 3.6% + 45
300 V 0.01 V 0.7% + 25 1.1% + 30 1.6% + 30 2.1% + 45 -
1000 V 0.1 V 0.7% + 25 1.1% + 30 1.6% + 30 - -
Notes for U1272A AC+DC voltage specifications:
1 Overload protection: 1000 Vrms. For millivolt measurements, 1000 Vrms for short circuits with <0.3 A current.
2 Input impedance: 10 MΩ (nominal) in parallel with <100 pF.
3 For 20 kHz to 100 kHz accuracy: Three counts of the LSD per kHz of additional error is to be added for frequencies >20 kHz
and signal inputs <10% of range.