Characteristics and Specifications 5
Electrical Specifications
U1271A/U1272A User’s Guide 165
AC+DC specifications for U1272A
Table 5-6 U1272A true rms AC+DC voltage specifications
Function Range Resolution
Accuracy
45 Hz to
65 Hz
20 Hz to
45 Hz
1 kHz to
5 kHz
5 kHz to
20 kHz
20 kHz to
100 kHz
[3]
65 Hz to
1 kHz
Voltage
30 mV 0.001 mV 0.7% + 40 0.8% + 45 1.1% + 45 1.1% + 60 3.6% + 60
300 mV 0.01 mV 0.7% + 25 0.8% + 30 1.1% + 30 1.1% + 45 3.6% + 45
3 V 0.0001 V 0.7% + 25 1.1% + 30 1.6% + 30 2.1% + 45 3.6% + 45
30 V 0.001 V 0.7% + 25 1.1% + 30 1.6% + 30 2.1% + 45 3.6% + 45
300 V 0.01 V 0.7% + 25 1.1% + 30 1.6% + 30 2.1% + 45 -
1000 V 0.1 V 0.7% + 25 1.1% + 30 1.6% + 30 - -
Notes for U1272A AC+DC voltage specifications:
1 Overload protection: 1000 Vrms. For millivolt measurements, 1000 Vrms for short circuits with <0.3 A current.
2 Input impedance: 10 MΩ (nominal) in parallel with <100 pF.
3 For 20 kHz to 100 kHz accuracy: Three counts of the LSD per kHz of additional error is to be added for frequencies >20 kHz
and signal inputs <10% of range.