3-2 Basic Operation PAS SERIES
The display that appears when the POWER switch is turned on for the
first time corresponds to factory default settings.
The unit stores the panel settings (excluding OUTPUT ON/OFF con-
dition)
immediately before the PO
WER switch is turned off. The pre-
vious panel settings are used the next time the POWER switch is
turned on.
For factory default settings, see below.
Factory default settings
Output voltage : 0 V
Output
current : M
aximum preset current
(105 % of the rated output current)
OVP (overvoltage protection): 110 % of the rated output voltage
OCP (overcurrent protection) : 110 % of the rated output current
CONFIG settings
CV control : 0 (Panel control)
CC control : 0 (Panel control)
Remote sensing : 0 (OFF)
PWR ON OUTPUT : 0 (OFF at startup)
Master-slave operation : 0 (MASTER/LOCAL)
EXT OUTPUT : 0 (HIGH=ON)
TERMN : 0 (OFF)
POWER switch trip : 0 (Enable)
3.2 Basic Operation
First, the procedure for setting the output and the protection function
trip point will be explained. Next, the procedure for using the unit as a
constant voltage or constant current power supply will be described.
For details on the display sections and switches, see Chapter6
"Names and Functions of Controls".