7.1 Specications
7.1.4 Detector Interface Unit
Item Description
Product name (model name) Detector Interface Unit (G-21)
Amount of connectable DR Detec-
tors
1
Power requirements
When the AC adapter is used:
Supplied from the DI Unit AC Adapter.
When the Power Supply Unit is used:
Supplied from the Power Supply Unit via the Ethernet cable.
Power supply when using the dedi-
cated AC adapter
AC 100 V-240 V ±10 %, single phase, 50/60 Hz
Power consumption when using
the dedicated AC adapter
2.62 A (With the DR Detector connected)
External dimensions
60(W)×160(D)×24(H) mm
Without mounting brackets: 60(W)×130(D)×22(H) mm
60 mm
160 mm
24 mm
Weight 0.3 kg
The dedicated AC adapter
specications
Product Name : SWITCHING POWER SUPPLY (Model Number. SINPRO ELECTRONICS
CO.,LTD. MPU64-108)
Dimensions : 76×146×43 mm (excluding cable)
Weight : 440 g
INPUT : AC100 V-240 V 1.62 A-0.72 A 47 Hz-63 Hz
OUTPUT : DC24 V max.2.62 A
Safety : IEC60601-1 Class I
7.1.5 Detector Interface Unit 2
Item Description
Product name (model name) Detector Interface Unit 2 (G-52)
Amount of connectable
DR Detectors
1
Power requirements Supplied from the Power Supply Unit via the Ethernet cable.
Power consumption 42.3 W (max) When charging Detector
External dimensions
85(W)×160(D)×24(H) mm (Excluding lugs)
Without mounting brackets: 85(W)×130(D)×22(H) mm
85 mm
160 mm
24 mm
Weight 500 g

Table of Contents