1.3 Safety precautions
Table 4 (support for IEC 60601-1-2:2007)
Recommended separation distance between portable and mobile RF communications equipment and the device
This device is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The
customer or the user of this device can help prevent electromagnetic interference by maintaining a minimum distance
between portable and mobile RF communications equipment (transmitters) and this device as recommended below,
according to the maximum output power of the communications equipment.
Rated maximum output
power of the transmitter
W
Separation distance according to frequency of transmitter
m
150 kHz to 80 MHz
d=[1.2] √P
80 MHz to 800 MHz
d=[1.2] √P
800 MHz to 2.5 GHz
d=[2.3] √P
0.01 0.12 0.12 0.23
0.1 0.38 0.38 0.73
1 1.2 1.2 2.3
10 3.8 3.8 7.3
100 12 12 23
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m)
can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power
rating of the transmitter in watts (W) according to the transmitter manufacturer.
[NOTE] At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
[NOTE] These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reection from structures, objects and people.
Table 5 (support for IEC 60601-1-2:2014)
Phenomenon
Basic EMC standard or test
method
Immunity test levels
Electrostatic discharge IEC 61000-4-2
±8 kV contact
±2 kV, ±4 kV, ±8 kV, ±15 kV air
Radiated RF EM elds IEC 61000-4-3
3 V/m
80 MHz - 2.7 GHz, 5.1 GHz - 5.8 GHz
80 % AM at 1 kHz
Proximity elds from RF
wireless communications
equipment
IEC 61000-4-3
See the following table "●Test specications for Enclosure port
immunity to RF wireless communications equipment"
Electrical fast transients /
bursts
IEC 61000-4-4
±2 kV input a.c. power port
±1 kV signal input/output parts port
100 kHz repetition frequency
Surges IEC 61000-4-5
±0.5 kV, ±1 kV line-to-line
±0.5 kV, ±1 kV, ±2 kV line-to-ground
Conducted desturbances
induced by RF elds
IEC 61000-4-6
3 V
0.15 MHz - 80 MHz
6 V in ISM bands between 0.15 MHz and 80 MHz
a
80 % AM at 1 kHz
Rated power frequency
magnetic elds
IEC 61000-4-8 30 A/m
Voltage dips IEC 61000-4-11
0 % U
T
; 0.5 cycle
At 0°, 45°, 90°, 135°, 180°, 225°, 270° and 315°
0 % U
T
; 1 cycle
and
70 % U
T
; 25/30 cycle
b
Single phase: at 0°
Voltage interruptions IEC 61000-4-11 0 % U
T
; 250/300 cycle
b
[NOTE] U
T
is the a.c. mains voltage prior to application of the test level.
a The ISM (industrial, scientic and medical) bands between 0.15 MHz and 80 MHz are 6.765 MHz to 6.795 MHz;
13.553 MHz to 13.567 MHz; 26.957 MHz to 27.283 MHz; and 40.66 MHz to 40.70 MHz.
b E.g. 10/12 means 10 periods at 50 Hz or 12 periods at 60 Hz.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Konica Minolta SKR 3000 and is the answer not in the manual?

Konica Minolta SKR 3000 Specifications

General IconGeneral
BrandKonica Minolta
ModelSKR 3000
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals