Krautkramer DM4E / DM4 / DM4 DL Issue 05, 01/01 4-9
Automatic probe zero adjustment Basics of operation
Accelerated setting
If, when setting, you have to span large ranges (e.g.
when setting the sound velocity), you can accelerate
this:
Press d or u and keep pressed.
The setting is accelerated.
4.6 Automatic probe zero
adjustment
With the DM4 you do not need to adjust the zero point
for the connected probe because the instrument auto-
matically carries out probe zero adjustment when the
probe is connected.
When using dialog probes, the probe zero adjustment
is continuously carried out even in the uncoupled state
(provided the contact face is clean). The active probe
zero adjustment is displayed by the following message
after the instrument is switched on:
When using non-dialog probes, the probe zero adjust-
ment is continuously carried out in the coupled state
during measurement.
With an active 2-point calibration, the automatic probe
zero adjustment is deactivated. Refer to Chapter 5.2.
Krautkramer DM4E / DM4 / DM4 DL Issue 05, 01/01 4-9
Automatic probe zero adjustment Basics of operation
Accelerated setting
If, when setting, you have to span large ranges (e.g.
when setting the sound velocity), you can accelerate
this:
Press d or u and keep pressed.
The setting is accelerated.
4.6 Automatic probe zero
adjustment
With the DM4 you do not need to adjust the zero point
for the connected probe because the instrument auto-
matically carries out probe zero adjustment when the
probe is connected.
When using dialog probes, the probe zero adjustment
is continuously carried out even in the uncoupled state
(provided the contact face is clean). The active probe
zero adjustment is displayed by the following message
after the instrument is switched on:
When using non-dialog probes, the probe zero adjust-
ment is continuously carried out in the coupled state
during measurement.
With an active 2-point calibration, the automatic probe
zero adjustment is deactivated. Refer to Chapter 5.2.