6-8 Issue 05, 01/01 Krautkramer DM4E / DM4 / DM4 DL
Operation of the Data Logger (DM4 DL only) Storing the measurement values
Inhibiting a memory location
You can inhibit a memory location and proceed to the
next location, for example when you are not able to
measure a particular point within a measurement se-
ries.
If needed, you can select the inhibited memory loca-
tion later in order to store a measurement value there.
The inhibited memory location is simply overwritten.
Operation:
You are in the mode THK, THK/SPEC or MIN and
have activated a file in which you wish to store your
values.
Uncouple the probe.
Press r.
As usual the memory location number will be briefly
displayed, e.g. L.003. Finally OBST (obstruct) will
appear.
The memory location is inhibited.
Simply carry out your next measurement.
The DM4 DL automatically stores the next measure-
ment value into the next memory location.
Selecting a memory location
You can individually select memory locations at which
the next measurement value is to be stored if, for ex-
ample, you overwrite an inhibited measurement value.
Operation:
You are in the mode THK, THK/SPEC or MIN:
Press y .
The active memory location number is displayed, e.g.
+L.003.
Whereby:
+L.NNN = memory location allocated
-L.NNN = memory location empty
6-8 Issue 05, 01/01 Krautkramer DM4E / DM4 / DM4 DL
Operation of the Data Logger (DM4 DL only) Storing the measurement values
Inhibiting a memory location
You can inhibit a memory location and proceed to the
next location, for example when you are not able to
measure a particular point within a measurement se-
ries.
If needed, you can select the inhibited memory loca-
tion later in order to store a measurement value there.
The inhibited memory location is simply overwritten.
Operation:
You are in the mode THK, THK/SPEC or MIN and
have activated a file in which you wish to store your
values.
Uncouple the probe.
Press r.
As usual the memory location number will be briefly
displayed, e.g. L.003. Finally OBST (obstruct) will
appear.
The memory location is inhibited.
Simply carry out your next measurement.
The DM4 DL automatically stores the next measure-
ment value into the next memory location.
Selecting a memory location
You can individually select memory locations at which
the next measurement value is to be stored if, for ex-
ample, you overwrite an inhibited measurement value.
Operation:
You are in the mode THK, THK/SPEC or MIN:
Press y .
The active memory location number is displayed, e.g.
+L.003.
Whereby:
+L.NNN = memory location allocated
-L.NNN = memory location empty