EasyManua.ls Logo

Krautkramer DM4E - 2-Point Calibration

Krautkramer DM4E
114 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Krautkramer DM4E / DM4 / DM4 DL Issue 05, 01/01 5-7
Calibrating the instrument Operation
2-point calibration
The 2-point calibration enables you to carry out an
accurate calibration. You should carry out this type of
calibration when you are required to carry out mea-
surements within a narrow thickness range.
You require a calibration block having two different
thicknesses (or two calibration blocks) which approxi-
mately correspond to the upper and lower limits of the
thickness range to be measured.
Operation:
Press m until the 2-point calibration function (2-PT)
is displayed. This function is switched of in the basic
setting.
Press c.
The CAL indication now flashes.
Switch the function on with d or u.
Press c again.
The following display appears:
Couple the probe to the step with the lowest thick-
ness on the calibration block. Make certain that the
coupling symbol appears and that the indication is
stable.
You can now decouple the probe because the
measurement value will remain.
Correct the displayed measurement value using d
u so that it corresponds with the actual thickness
of the calibration block.
Press c.
After this, calibrate the 2nd point (larger thickness).
Press c again.
Krautkramer DM4E / DM4 / DM4 DL Issue 05, 01/01 5-7
Calibrating the instrument Operation
2-point calibration
The 2-point calibration enables you to carry out an
accurate calibration. You should carry out this type of
calibration when you are required to carry out mea-
surements within a narrow thickness range.
You require a calibration block having two different
thicknesses (or two calibration blocks) which approxi-
mately correspond to the upper and lower limits of the
thickness range to be measured.
Operation:
Press m until the 2-point calibration function (2-PT)
is displayed. This function is switched of in the basic
setting.
Press c.
The CAL indication now flashes.
Switch the function on with d or u.
Press c again.
The following display appears:
Couple the probe to the step with the lowest thick-
ness on the calibration block. Make certain that the
coupling symbol appears and that the indication is
stable.
You can now decouple the probe because the
measurement value will remain.
Correct the displayed measurement value using d
u so that it corresponds with the actual thickness
of the calibration block.
Press c.
After this, calibrate the 2nd point (larger thickness).
Press c again.

Table of Contents

Other manuals for Krautkramer DM4E

Related product manuals