6-12 Issue 05, 01/01 Krautkramer DM4E / DM4 / DM4 DL
Operation of the Data Logger (DM4 DL only) Deleting stored measurement values
A message appears briefly indicating deletion of the
file (e.g. C.0.01). Then the file number appears.
A minus sign appears indicating that the file is empty,
e.g.:
The selected file is deleted.
With f return to the measurement mode or, using
d u , select a new file which you wish to delete.
Clearing the complete memory
You can delete the complete memory, i.e. all files with
all measurement values. For example, you must do
this when wishing to reconfigure the Data Logger.
A Attention:
When you delete a file, the data is irretrievably lost.
Save your data beforehand on an external data carrier
(refer to Chapter 7.2 Transferring data to a PC).
Operation:
Press f.
Press u or d until F.ALL appears. This indica-
tion appears at the beginning and at the end of the
data series, i.e. with u to the highest available file
number and with d to F.001.
Press r and keep pressed for about 3 seconds.
C.ALL is displayed.
Press f in order to return to the measurement
mode.
The complete memory is cleared. No measurement
values are stored until the Data Logger is
reconfigured, i.e. the number of files determined.
6-12 Issue 05, 01/01 Krautkramer DM4E / DM4 / DM4 DL
Operation of the Data Logger (DM4 DL only) Deleting stored measurement values
A message appears briefly indicating deletion of the
file (e.g. C.0.01). Then the file number appears.
A minus sign appears indicating that the file is empty,
e.g.:
The selected file is deleted.
With f return to the measurement mode or, using
d u , select a new file which you wish to delete.
Clearing the complete memory
You can delete the complete memory, i.e. all files with
all measurement values. For example, you must do
this when wishing to reconfigure the Data Logger.
A Attention:
When you delete a file, the data is irretrievably lost.
Save your data beforehand on an external data carrier
(refer to Chapter 7.2 Transferring data to a PC).
Operation:
Press f.
Press u or d until F.ALL appears. This indica-
tion appears at the beginning and at the end of the
data series, i.e. with u to the highest available file
number and with d to F.001.
Press r and keep pressed for about 3 seconds.
C.ALL is displayed.
Press f in order to return to the measurement
mode.
The complete memory is cleared. No measurement
values are stored until the Data Logger is
reconfigured, i.e. the number of files determined.