EasyManua.ls Logo

Krautkramer DM4E - Page 46

Krautkramer DM4E
114 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
5-8 Issue 05, 01/01 Krautkramer DM4E / DM4 / DM4 DL
Operation Calibrating the instrument
The following display appears:
Now calibrate the instrument to the larger thickness of
the calibration block. Proceed as described above.
After this, press c in order to exit the 2-PT function.
Check the calibration and, if required, carry out the
calibration one more time.
To do this, press m again first.
A Attention:
Check that only the status indicator symbol 2-PT is
displayed, no other status indicator symbol must ap-
pear on the display.
After this, press c to restart the calibration pro-
cess.
With the modes THK and MIN the 2-PT indication re-
mains at the top of the display giving an indication
about the current 2-point calibration of the instrument.
Automatic probe zero adjustment and 1-point calibra-
tion are now deactivated. Therefore, after a 2-point
calibration the PO indication for the automatic probe
zero adjustment no longer appears after the instru-
ment has been switched on.
Switching off 2-point calibration
Of course you can switch the 2-point calibration off
again in order to activate the automatic zero point
compensation and the 1-point calibration.
Press m until the 2-PT function is displayed.
Press c.
Set the function OFF with d or u.
Press c once again.
The 2-point calibration is now deactivated, the 2-PT
indication does not appear any more.
5-8 Issue 05, 01/01 Krautkramer DM4E / DM4 / DM4 DL
Operation Calibrating the instrument
The following display appears:
Now calibrate the instrument to the larger thickness of
the calibration block. Proceed as described above.
After this, press c in order to exit the 2-PT function.
Check the calibration and, if required, carry out the
calibration one more time.
To do this, press m again first.
A Attention:
Check that only the status indicator symbol 2-PT is
displayed, no other status indicator symbol must ap-
pear on the display.
After this, press c to restart the calibration pro-
cess.
With the modes THK and MIN the 2-PT indication re-
mains at the top of the display giving an indication
about the current 2-point calibration of the instrument.
Automatic probe zero adjustment and 1-point calibra-
tion are now deactivated. Therefore, after a 2-point
calibration the PO indication for the automatic probe
zero adjustment no longer appears after the instru-
ment has been switched on.
Switching off 2-point calibration
Of course you can switch the 2-point calibration off
again in order to activate the automatic zero point
compensation and the 1-point calibration.
Press m until the 2-PT function is displayed.
Press c.
Set the function OFF with d or u.
Press c once again.
The 2-point calibration is now deactivated, the 2-PT
indication does not appear any more.

Table of Contents

Other manuals for Krautkramer DM4E

Related product manuals