EasyManua.ls Logo

Krautkramer DM4E - Page 50

Krautkramer DM4E
114 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
5-12 Issue 05, 01/01 Krautkramer DM4E / DM4 / DM4 DL
Operation Thickness measurement
Carrying out the measurement
Make certain that the instrument is in the MIN mode
(if necessary press m).
Couple the probe onto the test object.
The coupling symbol appears at the same time as the
measured thickness value, e.g.:
Move the probe slowly over the range you wish to
measure.
The display refresh rate is 4 times per second; you
see the current measurment value.
Decouple the probe.
The smallest value in the measurement sequence is
displayed. The MIN indication on the display flashes
during the time-out.
For example:
If you couple the probe during the time-out the mea-
surement sequence will then be continued; the mea-
sured minimum thickness is retained.
When the time-out has finished the MIN indication
stops flashing. You can then start a new measurement
sequence.
5-12 Issue 05, 01/01 Krautkramer DM4E / DM4 / DM4 DL
Operation Thickness measurement
Carrying out the measurement
Make certain that the instrument is in the MIN mode
(if necessary press m).
Couple the probe onto the test object.
The coupling symbol appears at the same time as the
measured thickness value, e.g.:
Move the probe slowly over the range you wish to
measure.
The display refresh rate is 4 times per second; you
see the current measurment value.
Decouple the probe.
The smallest value in the measurement sequence is
displayed. The MIN indication on the display flashes
during the time-out.
For example:
If you couple the probe during the time-out the mea-
surement sequence will then be continued; the mea-
sured minimum thickness is retained.
When the time-out has finished the MIN indication
stops flashing. You can then start a new measurement
sequence.

Table of Contents

Other manuals for Krautkramer DM4E

Related product manuals