Operation
CoaDATA 501 – Operators Manual 2.1 Page 18
Quick
The curve for the PT calibration curve can be entered by typing in the 100
%-value and the factor for the slope of the calibration curve.
< quick >
• Press Enter to confirm selection.
<- 100 %
Example!
= 11.6 s
normal clotting time
factor = 54
Please use the value for factor as provided with the reagent package insert.
If you need to calculate the factor by yourself please refer to chapter 5.5.
If complete data has been entered for <curve> and <quick > the
calibration curve type selected last is active for con-versions in the
ready-to-measure mode. This is also valid for INR and ratio under
2nd conversion.
• Press Esc to access the measuring mode or Enter to add or verify
additional parameters.
none
If < none > was selected no conversion will be performed.
2nd conversion
< 2.conv >
If a calibration curve has been entered in menu 1st conversion or if <none>
has been selected, the 2nd conversion can be used for INR or ratio.
INR
• Press Enter and the following display appears:
< INR >
• Press Enter to type in the ISI-value.
ISI= 1.05
ATTENTION:
If <none> was selected under 1st conversion the 100 % sec
value, e.g. 100 % = 12,6 sec. will be requested.
Enter the ISI-value provided on the reagent package insert.
• Press Enter to confirm the entry.
NOTE