Operation
CoaDATA 501 – Operators Manual 2.1 Page 20
measurement
< measur >
• Press Enter
A display to type in the sample incubation time appears:
Incubation time
incubat.
= 60 s
• Enter the correct sample incubation time
• Press Enter to confirm the entry
A display which requests the entry of start reagent volume and lot number
appears.
Start reagent volume/
reagent lot number
<-Start Reagent
= 100 ul
• Enter the volume for the start reagent e.g. 100 µl
• Press Enter to confirm the entry
lot no.=
12345678
Example!
• Enter the reagent lot no.
• Press Enter to confirm the entry
<- select: ESC.
.. = work ENTER= more parameters
• Press Esc to access the measuring mode or Enter to add or verify
additional parameters.
cuvette remove detect.
< cuvdet >
• Press Enter-key to get the cuvette detection on the display.
< OFF >