Operation
CoaDATA 501 – Operators Manual 2.1 Page 35
print-out of method parameters
If you press 0 when the analyzer is in the measuring mode, a parameter
print-out for the selected method is generated.
print-outs PT
PT Documentation
Example: Conversions via a 4-point calibration curve in % and INR.
--- method store 1 ---
PT
actual date 02.12.99
cuv remove detection OFF
Automatic cuvette detection
mode DOUBLE
Double determination
coef.var = 5 %
Coefficient of variation
incubat = 60 s
Sample incubation time
start-reagent:
Lot 101xxx
Reagent lot no.
reagent = 100 ul
Reagent start volume
1st convers INTERPOLAT.
1. conversion, calib. curve / interpolation
2nd convers INR
2. conversion INR
ISI = 1.05
ISI-constant
100.0%= 11.6 s
Calibration curve points
50.0%= 17.7 s
25.0%= 29.9 s
10.0%= 66.6 s
-----------------------
results:
PT
Method
patient _____________
Patient name
02.12.99, 10:52:55
Date, Time
time 1 = 12.0 s
1st measured time
time 2 = 12.8 s
2nd measured time
Mean = 12.4 s
Mean of measured times
INR = 1.7
Conversion to INR
% = 88.4%
Conversion to PT %